Tubes
| The right
choice is decisive because an exactly and practically work requires
the right tube material. medorex offers these in different diameter
and wall thickness. |
| |
 |
|
- Silicone tubes with an excellent biocompatibility
- PharMed/Santopren, not toxic and not haemolytic, ideal
for cell- and tissue-cultures
- Tygon, the ideal tube for a lot of solvents
- Viton/Fluran F 5500, high Resitant against corrosive media
and high temperature range
- Norpren A60G, ideal for industrial use, heat and ozone
constant, abrasion resistant
|
| |
Tubes and their Application
Please choose the right tube material for your application
| Name |
Silicon |
PharMed/Santoprene |
Tygon
LFL |
| Advantages |
Hardly removing from softeners and other
additives smell neutral, not toxic, mould resistant, ideal for
low temperatures, waterproof and resistant to ozone, radiation,
sunlight
|
Can be welded, stuck and formed, very long
lifespan, low gas permeability, suitably for the medicine and
food area, impermeable for normal light and UV radiation.
|
Clear tube with a high life time, high chemical
compatibility, tasteless, good dielectric qualities, low gas
permeability, suitable for the medicine and Food section |
| Qualities |
Not suitably for concentrated solvents, oils, acids or thinned
potassium Hydroxide, relative high gas permeability |
Remove from additional chemicals is possible |
Not suitable for human blood and tissue part |
| Physical qualities |
Transparent, Polydimethylsiloxane with silica and Siliconoil
additives, excellent compression resistance
|
Polypropyleneelastomer, excellent tensile strength non-transparent,
beige
|
Flexible, soft, transparent clear |
| Temperature
range |
-50°C up to +230°C |
-50°C up to +135°C |
-50 up to +74°C |
| Application: |
|
|
|
| Acid |
Conditional |
good |
good |
| Base |
Conditional |
good |
good |
| Solvent |
not suitable |
not suitable |
not suitable |
| Pressure |
Satisfactory |
good |
good |
| Vacuum |
good |
good |
good |
| Viscous
media |
Satisfactory
till good |
excellent |
excellent |
| Sterile
media |
good |
good |
good |
| Complies
with the norms |
USP class VlFDA 21 CFR
177.2600 USDA standard |
USP class VlFDA 21 CFR 177.2600 (only Pharmed) 3 ARPA and
NSF
|
USP class
FDA |
| Sterilisation |
Gamma radiate or autoclavable in steam atmosphere, sterilisation
not with ethylene oxide |
Autoclavable, without ageing appearances |
Steam and ethylene oxide sterilization capable (max. 30 min.
at 121° degrees Celsius) |
| Permeability: |
|
|
|
| CO2 |
1200 |
1200 |
563 |
| H2 |
16170 x 10-11 |
772 x 10-11 |
362 x 10-11 |
| O2 |
200 |
200 |
124 |
| N2 |
80 |
80 |
67 |
| Name |
Viton
/ Fluran F-5500 |
Norprene |
| Advantages |
Resistant against chemicals, excellent resistance against
corrosive media, solvent and oils with high temperatures,
low gas permeability
|
Good resistance against acids and bases, long lifespan, low
gas permeability, can be stuck welded and formed, not oxidizing,
good dielectric qualities, doesn't age
|
| Qualities |
Restricted lifespan |
Remove from additional chemicals is possible |
| Physical qualities |
Fluorpolymerelastomer, non-transparent,
black |
Polypropylenelastomer, excellent tensile
strength, non-transparent, black |
| Temperature
range |
-40°C up to +200°C |
-60°C up to +130°C |
| Application: |
|
|
| acid |
excellent |
good |
| Base |
excellent |
good |
| Solvent |
good |
not suitable |
| Pressure |
good |
Satisfactory |
| Vacuum |
good |
good |
| Viscous media |
good |
excellent |
| Sterile
media |
Satisfactory |
Not recommendable |
| Complies
with the Norms |
no details |
no details |
| Sterilisation |
Steam and Ethylenoxide sterilization capable (max. 30 min.
at 121° degrees Celsius)
|
Not recommendable |
| Permeability: |
|
|
| CO2 |
38 |
1200 |
| H2 |
24 x 10-11 |
772 x 10-11 |
| O2 |
14 |
200 |
| N2 |
5 |
80 |
Manufacturer's indications, she should be used for